Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1396-2025

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior. Recall

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

FDA recall Class I

Ongoing

FDA lists this Class I recall for CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1396-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Feb 19, 2025
FDA report date
Apr 2, 2025
Quantity in commerce
11,246
Distribution
See key details
FDA cause
Device Design

What is being recalled?

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Affected model & UDI checker

Search the exact identifiers published for Z-1396-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CVC127020-1CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.Check source noticeNot specified
Not specifiedCVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.Check source notice00860005357710
FDA code information

CVC127020-1 UDI-DI: 00860005357710, L00018 Rev C

View the original FDA recall notice

Why was this product recalled?

Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.

What actions did Calyxo, Inc. take?

On 2/19/2025, correction notices were mailed to distributors and healthcare providers who were asked to do the following: 1) Notify all urologists utilizing the affected aspiration system in your institution that additional instructions are required. Please provide each of them with a copy of the additional instructions. 2) As another safety step, a firm representative will be contacting all users from your institution directly to offer training on these additional instructions. If you prefer to coordinate the training, please specifically state that preference in the acknowledgement. 3) Please acknowledge receipt of the notice, additional instructions, and updated training instructions MA00100.A by emailing qualitycontrol@calyxoinc.com, or calling 833-214-3354 and providing your name, title, institution and address.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology