Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1402-2025

Merge Cardio Recall

Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merge Cardio. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1402-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 19, 2025
FDA report date
Mar 26, 2025
Quantity in commerce
253
Distribution
See key details
FDA cause
Software design

What is being recalled?

Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.

Affected model & UDI checker

Search the exact identifiers published for Z-1402-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMerge CardioCheck source notice00842000101123
Not specifiedMerge CardioCheck source notice00842000100928
Not specifiedMerge CardioCheck source notice00842000100065
Not specifiedMerge CardioCheck source notice00842000101109
FDA code information

Lot Code: Software versions and UDIs 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206 12.1, UDI: (01)00842000100928(10)12.1(11)210122; 12.1.1, UDI: (01)00842000100928(10)12.1.1(11)211220; 12.2 UDI: (01)00842000100065(10)12.2.0(11)220713; 12.3, UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.1 UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.2 UDI: (01)00842000101109(10)12.3.2(11)230612; 12.3.3 UDI: (01)00842000101109(10)12.3.3(11)230913; 12.4 UDI: (01)00842000101123(10)12.4.0(11)231220; 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522; 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206

View the original FDA recall notice

Why was this product recalled?

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

What actions did Merge Healthcare, Inc. take?

Merge Healthcare issued an Urgent Field Safety Notice to it's consignees on 02/19/2025 via email. The notice explained the issue, risk to patients, provided a workaround to immediately address the issue, and recommended clinicians check measurements in the final report to ensure that they accurately reflect the patient's condition pending software update. Required Actions: Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Schedule an update to implement the configuration change on your system. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge. Please forward this notice to any individuals within your organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The firm identified a missing software version (11.0.2) and issued the corrected notice on 03/03/2025.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology