Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627 Recall
Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627
FDA recall Class II
Ongoing
FDA lists this Class II recall for Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1403-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 3, 2025
- FDA report date
- Mar 26, 2025
- Quantity in commerce
- 90 US; 136 OUS
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627
Affected model & UDI checker
Search the exact identifiers published for Z-1403-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627 | Check source notice | 00887868267550 |
| Not specified | Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627 | Check source notice | 00887868267567 |
| Not specified | Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627 | Check source notice | 00887868267574 |
| Not specified | Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627 | Check source notice | 00887868267741 |
FDA code information
Item No. 113605, UDI-DI 00887868267550 Lots 66173214, 66173215; Item No. 113606, UDI-DI 00887868267567 Lots 66173216, 66173218, 66173219; Item No. 113607, UDI-DI 00887868267574 Lots 66173220, 66173221; Item No. 113627, UDI-DI 00887868267741 Lots 65901697, 65909480, 65915809
Why was this product recalled?
Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
What actions did ZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAY take?
On March 3, 2025, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were informed of the excess metal on the inner taper of the stem due to a chipped manufacturing tool. Customers were asked to notify all affected persons and assist their Zimmer Biomet sales representative in quarantining all affected product on hand. The products will be removed from your facility.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1403-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology