Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1407-2025

Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200. Recall

Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1407-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 11, 2025
FDA report date
Mar 26, 2025
Quantity in commerce
104 units (69 US, 35 OUS)
Distribution
See key details
FDA cause
Software change control

What is being recalled?

Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.

Affected model & UDI checker

Search the exact identifiers published for Z-1407-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
7.4.001Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.Check source noticeNot specified
7.4.010Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.Check source noticeNot specified
7.4.100Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.Check source noticeNot specified
7.4.110Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.Check source noticeNot specified
7.4.200Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.Check source noticeNot specified
Not specifiedSynapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.Check source notice00854904006008
FDA code information

Model numbers: Synapse PACS 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200 Software Version Numbers: Synapse PACS 7.4.000, UDI: (01)00854904006008(10)0704000; Synapse PACS 7.4.001, UDI: (01)00854904006008(10)0704001; Synapse PACS 7.4.010, UDI: (01)00854904006008(10)0704010; Synapse PACS 7.4.100, UDI: (01)00854904006008(10)0704100; Synapse PACS 7.4.110, UDI: (01)00854904006008(10)0704110; Synapse PACS 7.4.200, UDI: (01)00854904006008(10)0704200.

View the original FDA recall notice

Why was this product recalled?

The incorrect computed patient age is showing in VX for patients less than 3 months old.

What actions did FUJIFILM Healthcare Americas Corporation take?

On February 11, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. ACTIONS TO BE TAKEN BY THE COMPANY Fujifilm will issue corrections for all impacted 7.4.x versions. The following patch releases will be made available: - 7.4.020 for 7.4.0x versions (immediately available) - 7.4.120 for 7.4.1x versions (available February 28, 2025) - 7.4.210 for 7.4.2x versions (available February 28,2025) ACTIONS THAT CAN BE TAKEN BY THE END-USER IMMEDIATELY Regarding this issue, Synapse PACS VX can safely be used by the end-user as long as; 1] the end-user is aware of the issue and, 2] takes appropriate actions when rendering a diagnosis for patients under 18 years of age. The following workarounds can be used: 1) The image acquisition date and date of birth are displayed on the Image. The end-user can use the dates to verify the age or, 2) Open the Series Viewer by double clicking on the thumbnail displayed on top the Synapse VX Image viewing window or the worklist. The Series Viewer Window will render the image on the server using the Server-side rendering (Zero Viewer component) which will calculate the age correctly. Alternatively, the user can switch to the Synapse PACS Zero Viewer/Thin Viewer (Server-side Rendering). The Synapse PACS Zero Viewer/Thin Viewer calculates the age correctly. ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1) Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 2) Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to schedule a patch update. Fujifilm regrets any inconvenience this action may cause and appreciates your understanding and cooperation. Please contact our field action coordinator, Jay Wigley, via email at hcusquality@fujifilm.com, if you have any further questions regarding this field action. Please contact Technical Assistance Center (TAC) at 1-888-385-4622 or 1-888-FU

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.