Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1411-2025

Baxter Floor mounting column TruSystem 7500, Product Code 1717021 Recall

Baxter Floor mounting column TruSystem 7500, Product Code 1717021

FDA recall Class II

Ongoing

FDA lists this Class II recall for Baxter Floor mounting column TruSystem 7500, Product Code 1717021. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1411-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 25, 2025
FDA report date
Apr 2, 2025
Quantity in commerce
112 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Baxter Floor mounting column TruSystem 7500, Product Code 1717021

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UID/DI 00887761974401, All serial numbers manufactured until 11/14/2024.

View the original FDA recall notice

Why was this product recalled?

There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.

What actions did Baxter Healthcare Corporation take?

Baxter issued an UREGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/25/2025 by letter sent via traceable means. The notice explained the issue, hazard involved, and requested the following actions be taken: "Actions to be Taken by Customers 1. Baxter will contact you to install a software upgrade in the impacted surgical table systems. 2. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply form instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 3. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 4. Please forward a copy of this communication to the Medical Director/Office Manager, Director of Pharmacy, Medication Safety Officer, Purchasing Director/Head of Central Supply and any other departments within your institution who use the affected product. 5. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal." "For general questions regarding this communication, please contact Baxter Technical Support at 800-445-3720 (select option 2, then option 7) between 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or email Baxter at HRC_Trumpf_Tech_Support@baxter.com."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology