namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit Recall
namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit
FDA recall Class II
Ongoing
FDA lists this Class II recall for namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1428-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 12, 2025
- FDA report date
- Apr 2, 2025
- Quantity in commerce
- 4400 units
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit
Affected model & UDI checker
Search the exact identifiers published for Z-1428-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 80641722 | namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit | Check source notice | Not specified |
| Not specified | namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit | Check source notice | 10193489067507 |
| Not specified | namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical convenience kit | Check source notice | 20193489067504 |
FDA code information
UDI/DI 10193489067507 (each), 20193489067504 (case), Lot number 0000140528
Why was this product recalled?
The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
What actions did MEDLINE INDUSTRIES, LP - Northfield take?
Medline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medline will provide labeling that states the following: "ATTENTION This kit contains a recalled item Medline Pressure Monitoring Lines VPN: 24614303 DO NOT USE THIS COMPONENT The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*" Consignees were directed to notify their customers of these actions.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1428-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology