Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1428-2026

Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Recall

Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/Catalog Number: 380730 Used for minimally invasive surgery.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Da Vinci 5 ASSY, DV5 CONSOLE, IS5000. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1428-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 17, 2025
FDA report date
Feb 25, 2026
Quantity in commerce
12
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/Catalog Number: 380730 Used for minimally invasive surgery.

Affected model & UDI checker

Search the exact identifiers published for Z-1428-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
380730Da Vinci 5 ASSY, DV5 CONSOLE, IS5000Check source noticeNot specified
Not specifiedDa Vinci 5 ASSY, DV5 CONSOLE, IS5000Check source notice00886874119747
FDA code information

UDI-DI: 00886874119747 GTIN: 0886874119747 with software version P1.2.1

View the original FDA recall notice

Why was this product recalled?

An error event following the release of a software version for the surgical system that can result in loss of user interface content on an external monitor or tower monitor.

What actions did Intuitive Surgical, Inc. take?

On 11/17/2025, the firm emailed a "New Customer Communication" to customers informing them that Intuitive has become aware, following the distribution of P1.2.1 software version for dV5 Systems between October 17, 2025 and October 24, 2025, of an increase in 48306 error events observed on Da Vinci 5 Surgical Systems in dual console configuration, causing User Interface (UI) application crashes and loss of interaction at the affected node. Customers are instructed to: Continue Safe Use: If error 48306 occurs, follow the on-screen instructions and select Retry to recover. Affected systems may continue to be used following the recovery of error 48306 with the awareness of potential loss of non-essential UI features as outlined below: " User interface may not be displayed on connected external monitor live endoscopic video is still displayed " User interface may not be displayed on the dV5 Tower monitor - live endoscopic video is still displayed " System may be unable to activate Tile/Overlay layout user is unable to stream video input from external sources into the Console viewer. If UI streaming remains unavailable, restart the system during the procedure to restore full functionality. Guidance is available in Appendix B, Faults that Require Restarting , of Da Vinci 5 User Manual PN 555500-05. Intuitive Surgical is developing a corrective software patch (version P1.2.2). When ready, customers will be contacted to schedule installation. Please take the additional following actions: 1. Display this communication with systems; ensure it is in a place likely to be seen/viewed by operators. 2. Retain a copy of this letter and the acknowledgement form for your files. For questions - contact Intuitive Representative or contact Intuitive Customer Service at: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology