Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1430-2026

Brand Name: Hitachi Proton Beam Therapy REF: PROBEAT-CR Recall

Brand Name: Hitachi Proton Beam Therapy REF: PROBEAT-CR

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Hitachi Proton Beam Therapy REF: PROBEAT-CR. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1430-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 6, 2026
FDA report date
Mar 4, 2026
Quantity in commerce
1 unit
Distribution
See key details
FDA cause
Software design

What is being recalled?

Brand Name: Hitachi Proton Beam Therapy REF: PROBEAT-CR

Affected model & UDI checker

Search the exact identifiers published for Z-1430-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedBrand Name: Hitachi Proton Beam Therapy REF: PROBEAT-CRCheck source notice04560333350006
FDA code information

Serial Number: CRUS01/ UDI: (01)04560333350006(21)CRUS01(11)190920

View the original FDA recall notice

Why was this product recalled?

Software anomaly in the patient positioning system may result in positional discrepancy.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha take?

Between the dates of February 5-11, 2026, Hitachi, Ltd, issued an "URGENT: MEDICAL DEVICE RECALL" notification dated January 21, 2026, via Service Team Manager to affected consignees. Hitachi is asking consignees to take the following actions: 1. Until the PPS software has been updated with the corrected version, users are instructed to follow the System Operation instruction in the document. 2. Users are also asked to post the document in a prominent space and sign its cover. 3. Complete and return the acknowledgment and receipt form. Action by the firm: Hitachi plans to update the PPS software to a corrected version that prevents the mismatch between the commanded error correction parameter and the internal variable. starting in late March 2026 and the users will be contacted after the update. If you have any questions, please call QA Manager at Tel: (408) 986-6300. Monday through Friday, 8:00 AM to 5:00 PM, Central Time.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology