Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1473-2026

Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1 Recall

Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1

FDA recall Class II

Ongoing

FDA lists this Class II recall for Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1473-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 13, 2026
FDA report date
Mar 4, 2026
Quantity in commerce
939 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1

Affected model & UDI checker

Search the exact identifiers published for Z-1473-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedOctopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1Check source notice00763000543693
FDA code information

UDI-DI: 00763000543693; Lot Number: 0232837110, Additional lots as of 5/22/26: 0231547054 0231591827 0231758880 0231758898 0231758932 0231792414 0231792421 0231912903 0232670901 0232709219 0232709254 0232744928 0232784925 0232784989 0232805458

View the original FDA recall notice

Why was this product recalled?

During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

What actions did MEDTRONIC PERFUSION SYSTEMS take?

An URGENT: MEDICAL DEVICE RECALL notification letter dated January 2026 was hand delivered to customers. Actions: Medtronic requests that you take the following actions: " Review your inventory for listed lot number. " If you have inventory from the listed lot number immediately quarantine and return it to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of this unit. Your Medtronic sales representative can assist you in the return of unit as necessary. " Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have the affected product in your possession. " Please maintain a copy of this communication in your records. If you have any questions regarding this communication, please contact your Medtronic Field Representative. An URGENT: MEDICAL DEVICE RECALL notification letter dated 5/12/26 was sent to customers. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for lot numbers listed in Attachment A. " If you have unused inventory from the listed lot numbers, immediately quarantine and return the products to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of this unit. Your Medtronic sales representative can assist you in the return of unit as necessary. " Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have the affected product in your possession. " Please share this notification with others in your organization as appropriate. If product listed above has been forwarded to another facility, please notify the facility of this Medtronic Urgent Medical Device Recall. " Please maintain a copy of this communication in your records. If you

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology