Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL Recall
Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL
FDA recall Class II
Ongoing
FDA lists this Class II recall for Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1474-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 25, 2025
- FDA report date
- Apr 9, 2025
- Quantity in commerce
- Not reported
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL
Affected model & UDI checker
Search the exact identifiers published for Z-1474-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 5400-020-004 | Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL | Check source notice | Not specified |
| 5400-020-004ITL | Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL | Check source notice | Not specified |
| Not specified | Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL | Check source notice | 07613154340811 |
| Not specified | Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL | Check source notice | 07613327359459 |
FDA code information
5400-020-004 UDI-DI: 07613154340811, Lot Numbers: 2022100314 2023013132 2023042514 2023081512 2023110114 2024011514 2024032512 2024082612 2022102514 2023021012 2023050216 2023082811 2023110614 2024011814 2024042914 2023011332 2023021512 2023060214 2023090514 2023111616 2024021213 2024052812 2023012632 2023030814 2023062214 2023091813 2023112112 2024030412 2024073012 ; 5400-020-004ITL UDI-DI: 07613327359459, Lot Numbers: 2022102012 2023011714 2023011714 2023051614 2023081812 2023110914 2024012614 2024030716 2022111514 2023012014 2023012014 2023052314 2023083116 2023112414 2024020714 2024031214 2023010914 2023020214 2023020214 2023062733 2023100914 2023112912 2024021514 2024031814 2023011214 2023051112 2023051112 2023080112 2023101214 2024010912 2024022014 2024032816 2024042314 2024051314 2024052212 2024060512 2024061114
Why was this product recalled?
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Stryker Corporation take?
An Urgent Medical Device Recall notification letter dated 2/24/25 was sent to customers. Actions to be taken 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. Instruments.recalls@stryker.com. RESPONSE IS REQUIRED.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1474-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology