Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1474-2026

Campy CVA Medium 100/PK, R01272 Recall

Campy CVA Medium 100/PK, R01272

FDA recall Class II

Ongoing

FDA lists this Class II recall for Campy CVA Medium 100/PK, R01272. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1474-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 5, 2026
FDA report date
Mar 4, 2026
Quantity in commerce
97
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Campy CVA Medium 100/PK, R01272

Affected model & UDI checker

Search the exact identifiers published for Z-1474-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedCampy CVA Medium 100/PK, R01272Check source notice848838001055
FDA code information

UDI-DI 848838001055 Lot R01272 lot 339135 Expired January 12, 2026

View the original FDA recall notice

Why was this product recalled?

Customer complaints report low to no recovery of Campylobacter Jejuni ATCC 33291 on identified lot of medium.

What actions did Remel, Inc take?

On February 5, 2026, the firm notified affected customers via Urgent Medical Device Recall letter. Customers were instructed to review test results and destroy any remaining inventory of the affected lot.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.