GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045. Recall
GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.
FDA recall Class II
Ongoing
FDA lists this Class II recall for GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1487-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 19, 2025
- FDA report date
- Apr 9, 2025
- Quantity in commerce
- 4 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.
Affected model & UDI checker
Search the exact identifiers published for Z-1487-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045. | Check source notice | 00840682103800 |
FDA code information
Model/catalog/code # 2049587-015, Installed Product ID 100707-1-Centricity Universal Viewer-00558649, UDI UDI Not Required, System ID PACS-IW-126; Model/catalog/code # 2066908-086, Installed Product ID 9842-4-Centricity Universal Viewer-19386293, UDI UDI Not Required, System ID GA001PACTRL01; Model/catalog/code # 2066908-136, Installed Product ID ZA2533-Centricity Universal Viewer-00116577, UDI UDI Not Required, System ID ZA2533PAC1; Model/catalog/code # 2104867-045, Installed Product ID X9003062 - HC3062-Centricity Universal Viewer-01462397, UDI (01)00840682103800(10)6.0SP50412094097001D, System ID HC3062UV01.
Why was this product recalled?
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did GE Healthcare take?
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 by letter delivered using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification. Please ensure you have implemented the actions below: 1: Ensure your hospital network is secure and 2: Outside of the hospital network implement additional safeguards such as a VPN. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.85480@gehealthcare.com." "If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative." Per status report 05/23/2025, the firm send a follow-up notice dated 05/13/2025 providing an installation manual addendum that provides instructions for changing passwords.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1487-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology