Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1487-2025

GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045. Recall

GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1487-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 19, 2025
FDA report date
Apr 9, 2025
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.

Affected model & UDI checker

Search the exact identifiers published for Z-1487-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.Check source notice00840682103800
FDA code information

Model/catalog/code # 2049587-015, Installed Product ID 100707-1-Centricity Universal Viewer-00558649, UDI UDI Not Required, System ID PACS-IW-126; Model/catalog/code # 2066908-086, Installed Product ID 9842-4-Centricity Universal Viewer-19386293, UDI UDI Not Required, System ID GA001PACTRL01; Model/catalog/code # 2066908-136, Installed Product ID ZA2533-Centricity Universal Viewer-00116577, UDI UDI Not Required, System ID ZA2533PAC1; Model/catalog/code # 2104867-045, Installed Product ID X9003062 - HC3062-Centricity Universal Viewer-01462397, UDI (01)00840682103800(10)6.0SP50412094097001D, System ID HC3062UV01.

View the original FDA recall notice

Why was this product recalled?

There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Healthcare take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 by letter delivered using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification. Please ensure you have implemented the actions below: 1: Ensure your hospital network is secure and 2: Outside of the hospital network implement additional safeguards such as a VPN. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.85480@gehealthcare.com." "If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative." Per status report 05/23/2025, the firm send a follow-up notice dated 05/13/2025 providing an installation manual addendum that provides instructions for changing passwords.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology