Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1490-2026

Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038 Recall

Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038

FDA recall Class II

Completed

FDA lists this Class II recall for Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1490-2026
Recall classification
Class II
Enforcement status
Completed
Initiation date
Mar 14, 2025
FDA report date
Mar 11, 2026
Quantity in commerce
5 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038

Affected model & UDI checker

Search the exact identifiers published for Z-1490-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
120000038Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038Check source noticeNot specified
Not specifiedEsaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038Check source notice0805604453891
FDA code information

UDI 0805604453891, Serial Numbers: 241700533C, 241700529C, 241700607C, 241700536C, 241700613C.

View the original FDA recall notice

Why was this product recalled?

A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did ESAOTE S.P.A. take?

Esaote issued a Field Safety Notice to its consignees on 3/14/2025 via FedEx. The notice explained the problem with the device, potential risk of use and requested the following: "Actions to be performed by the user " If the operator detects the issue during the inspection, they must NOT use the probe and should contact Esaote's technical support. " Please distribute this Safety Notice to all personnel involved within your organization and to all users of the device. " If the device is no longer in your possession, please provide Esaote with the new owner's details or inform us about the current status of the device. As a precautionary measure, and to complete the root cause analysis while improving our processes and customer service, Esaote will proceed with a voluntary precautionary recall. Our Technical Support Service will soon contact customers who own the affected units to arrange for the replacement of the probe."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology