Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038 Recall
Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038
FDA recall Class II
Completed
FDA lists this Class II recall for Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1490-2026
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- Mar 14, 2025
- FDA report date
- Mar 11, 2026
- Quantity in commerce
- 5 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038
Affected model & UDI checker
Search the exact identifiers published for Z-1490-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 120000038 | Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038 | Check source notice | Not specified |
| Not specified | Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038 | Check source notice | 0805604453891 |
FDA code information
UDI 0805604453891, Serial Numbers: 241700533C, 241700529C, 241700607C, 241700536C, 241700613C.
Why was this product recalled?
A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did ESAOTE S.P.A. take?
Esaote issued a Field Safety Notice to its consignees on 3/14/2025 via FedEx. The notice explained the problem with the device, potential risk of use and requested the following: "Actions to be performed by the user " If the operator detects the issue during the inspection, they must NOT use the probe and should contact Esaote's technical support. " Please distribute this Safety Notice to all personnel involved within your organization and to all users of the device. " If the device is no longer in your possession, please provide Esaote with the new owner's details or inform us about the current status of the device. As a precautionary measure, and to complete the root cause analysis while improving our processes and customer service, Esaote will proceed with a voluntary precautionary recall. Our Technical Support Service will soon contact customers who own the affected units to arrange for the replacement of the probe."
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1490-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology