Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1497-2026

IBA Proton Therapy System - PROTEUS 235 Recall

IBA Proton Therapy System - PROTEUS 235

FDA recall Class II

Ongoing

FDA lists this Class II recall for IBA Proton Therapy System - PROTEUS 235. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1497-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 3, 2026
FDA report date
Mar 18, 2026
Quantity in commerce
4 units (1 US, 3 OUS)
Distribution
See key details
FDA cause
Software design

What is being recalled?

IBA Proton Therapy System - PROTEUS 235

Affected model & UDI checker

Search the exact identifiers published for Z-1497-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedIBA Proton Therapy System - PROTEUS 235Check source notice05404013801138
FDA code information

PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)

View the original FDA recall notice

Why was this product recalled?

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

What actions did Ion Beam Applications S.A. take?

On February 3, 2026 Urgent Medical Device Correction letters were sent to customers. Letter provides awareness of the Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. On-site modification of the devices will be performed by IBA with a TSS Configuration update. The update will be deployed at the impacted sites by June 2026.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology