FDA RECALL RECORD · Z-1500-2025

Phoroptor VRx Digital Refraction System Model Numbers: 16241 Recall

Phoroptor VRx Digital Refraction System Model Numbers: 16241

FDA recall Class II

Ongoing

FDA lists this Class II recall for Phoroptor VRx Digital Refraction System Model Numbers: 16241. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1500-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 28, 2025
FDA report date
Apr 9, 2025
Quantity in commerce
122 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Phoroptor VRx Digital Refraction System Model Numbers: 16241

Affected model & UDI checker

Search the exact identifiers published for Z-1500-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
16241Phoroptor VRx Digital Refraction System Model Numbers: 16241Check source noticeNot specified
FDA code information

Model Numbers: (1) 16241; UDI-DI: (1) 00812559011716; Serial Numbers: TBD; Note: The majority of units within the scope of the planned recall are older devices which do not have a UDI on the label.

View the original FDA recall notice

Why was this product recalled?

The head of the phoropter head could come loose and potentially detach due to a default in the assembly.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did REICHERT INC. take?

On February 28, 2025 "URGENT MEDICAL DEVICE RECALL" letters were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS IMMEDIATELY. 1. Inspect your Phoroptor head with reference to Figure 1 for looseness or rocking and Figure 2 for an excessive gap between components. Figure 3 shows the normal condition. 2. If your unit exhibits looseness or rocking or if there is an excessive gap between components, it is possible that your unit has the defect condition. Please remove it from service immediately and contact Reichert Technical Support at 1-888-849-8955 or at https://www.reichert.com/en/support/rma-request so that the unit can be inspected and, if necessary, repaired by Reichert. The repair will be completed at no cost to you. 3. If your unit shows the normal condition as in Figure 3, please install the Support Bracket when you receive it. It is essential that you install the Support Bracket for the safety of patients and operators even if the components appear tight and securely connected. Without the bracket, there could be eventual separation of the components. The Support Bracket prevents this possibility. The internal investigation into this issue has indicated the need for these corrections and resulted in a plan that includes actions designed to prevent a recurrence of this situation. Reichert prides itself on the quality of devices we manufacture. Please share this letter with all the relevant personnel in your organization. Please contact Reichert Technical Support with any questions. They can be reached Monday through Friday, 8:00AM to 5:00PM Eastern Time at 1-888-849-8955 or at https://www.reichert.com/en/support/support-and-service-inquiry.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.