Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1523-2026

Endo-Model Replacement Plateau; Item Number: 15-0027/16; Recall

Endo-Model Replacement Plateau; Item Number: 15-0027/16;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Endo-Model Replacement Plateau; Item Number: 15-0027/16;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1523-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 12, 2026
FDA report date
Mar 18, 2026
Quantity in commerce
3 units (OUS only)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Endo-Model Replacement Plateau; Item Number: 15-0027/16;

Affected model & UDI checker

Search the exact identifiers published for Z-1523-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedEndo-Model Replacement Plateau; Item Number: 15-0027/16;Check source notice04026575432134
FDA code information

Item Number: 15-0027/16; UDI-DI: 04026575432134; Serial/Lot Number: 2550412, 2550413, 2549228;

View the original FDA recall notice

Why was this product recalled?

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Waldemar Link GmbH & Co. KG (Mfg Site) take?

On or about January 12, 2026 URGENT FIELD SAFETY NOTICE Product Recall letters were sent to customers. Actions to be taken: Review inventory to identify and quarantine any affected product. Return any unused product to the manufacturer. Follow patient management recommendations, if implanted. A custom-made replacement set can be requested. If an affected product is noticed during implantation, another plateau screw from a replacement set or different LINK knee prosthesis package can be used. If this situation occurs, a complaint form must be completed and sent to Complaint@link-ortho.com. The replacement of the affected products will begin as soon as replacement products are available. Replacement will not incur any costs to you. Should you have any questions on acquiring replacements for forthcoming surgeries, please contact your local sales representative or customer service for Link products.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology