PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Recall
PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).
FDA recall Class II
Ongoing
FDA lists this Class II recall for PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1544-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 21, 2026
- FDA report date
- Mar 18, 2026
- Quantity in commerce
- 16,499 electrodes (US/OUS: 8327/8172)
- Distribution
- See key details
- FDA cause
- Unknown/Undetermined by firm
What is being recalled?
PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).
Affected model & UDI checker
Search the exact identifiers published for Z-1544-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 11101-000016 | PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 | Check source notice | Not specified |
| 11101-000017 | PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 | Check source notice | Not specified |
| Not specified | PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 | Check source notice | 00721902629013 |
| Not specified | PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 | Check source notice | 00721902682483 |
FDA code information
Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Lot Numbers: 315303, 315639, 315858, 316007, 320529, 318931
Why was this product recalled?
Due to pediatric defibrillator electrode delamination
What actions did Physio-Control, Inc. take?
On 01/21/2026, the firm sent via certified mail an "Urgent: Medical Device Recall" letter to customer informing them that Stryker was notified of complaints that the pediatric electrodes were experiencing gel delamination and could lead to a noisy or distorted ECG waveform and in addition may deliver energy to a patient without the electrode's gel barrier may result in burns. Customers are instructed to: 1. Immediately check internal inventory to locate the affected product as listed above. 2. Stop use and segregate and quarantine the affected product. 3. On the business reply form, indicate if an affected electrode is in their possession. 4. Return the business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 5. Upon receipt of the completed business reply form, Stryker will contact you to arrange for replacement of the electrode. 6. A return material authorization (RMA) will be provided when a replacement electrode is sent to return the affected electrodes back to Stryker. 7. Maintain awareness of this communication internally until all required actions have been completed within their facility. Ensure this letter is kept with the affected device until the correction has been completed. 8. Inform Stryker if any of the subject electrodes have been distributed to other organizations. If so, provide contact details, so Stryker can inform the recipients appropriately. For Questions - contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at medtechsup@stryker.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1544-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology