EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; Recall
EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;
FDA recall Class II
Ongoing
FDA lists this Class II recall for EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1551-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 10, 2026
- FDA report date
- Mar 18, 2026
- Quantity in commerce
- 806 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;
Affected model & UDI checker
Search the exact identifiers published for Z-1551-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; | Check source notice | 08714729880394 |
| Not specified | EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; | Check source notice | 08714729880387 |
FDA code information
Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; Outer Box UDI: 08714729880394; Inner Box UDI: 08714729880387; Outer Box Lot#s: 37435463, 37522541, 37528545, 37546847, 37546850, 37546849, 37550500, 37592036, 37732292; Inner Box Lot#s: 37435462, 37514809, 37525231, 37536563, 37536564, 37536562, 37546840, 37583642, 37731897;
Why was this product recalled?
Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did BOSTON SCIENTIFIC CORPORATION take?
On February 10, 2026 Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: " Do NOT use affected product (Attachment 1), remove affected product from your facility s inventory, and segregate the units in a secure place until they can be returned to Boston Scientific. " Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. " Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. " Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions. " Return affected product. This removal affects only the products and lots affected. No other material numbers or lots are impacted by this removal. If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1551-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology