FDA RECALL RECORD · Z-1553-2025

Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System. Recall

Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1553-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 10, 2025
FDA report date
Apr 23, 2025
Quantity in commerce
83 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

Affected model & UDI checker

Search the exact identifiers published for Z-1553-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
712203Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.Check source noticeNot specified
Not specifiedPhilips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.Check source notice00884838095205
FDA code information

REF 712203; UDI: (01)00884838095205(21); Serial No. 220025, 210023, 220171, 230103, 230152.

View the original FDA recall notice

Why was this product recalled?

Potential for collimator to fall as a result of incorrect installation.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America Llc take?

Consignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an inspection of devices by field service engineers. Engineers may also replace the collimator, if necessary. Consignees with questions are to call Philips Customer Care Solutions Center at 1-800-722-9377 from Monday through Friday (8:00 AM to 8:00 PM EST).

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.