Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System. Recall
Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1553-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 10, 2025
- FDA report date
- Apr 23, 2025
- Quantity in commerce
- 83 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.
Affected model & UDI checker
Search the exact identifiers published for Z-1553-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 712203 | Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System. | Check source notice | Not specified |
| Not specified | Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System. | Check source notice | 00884838095205 |
FDA code information
REF 712203; UDI: (01)00884838095205(21); Serial No. 220025, 210023, 220171, 230103, 230152.
Why was this product recalled?
Potential for collimator to fall as a result of incorrect installation.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips North America Llc take?
Consignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an inspection of devices by field service engineers. Engineers may also replace the collimator, if necessary. Consignees with questions are to call Philips Customer Care Solutions Center at 1-800-722-9377 from Monday through Friday (8:00 AM to 8:00 PM EST).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1553-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology