Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1557-2026

HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS). Recall

HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS).

FDA recall Class II

Ongoing

FDA lists this Class II recall for HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1557-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 30, 2026
FDA report date
Mar 18, 2026
Quantity in commerce
78 systems
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS).

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: CFN: VSLBASE03P / VITALSYNCSW03 UDI-DI (GTIN); 10884521844933 Serial Number: Vital Sync 3.3.1 CFN: VSLBASE03P / VITALSYNCSW03 UDI-DI (GTIN): 10884521844940 Serial Number: Vital Sync 3.4.0 CFN: VSLBASE03P / VITALSYNCSW03 (OUS only) UDI-DI (GTIN): 10884521847569 Serial Number: Vital Sync 3.4.1 CFN: VSLBASE03P / VITALSYNCSW03 UDI-DI (GTIN): 10884521856127 Serial Number: Vital Sync 3.4.2 CFN: VSLBASE04P / VITALSYNCSW04 UDI-DI (GTIN): 10884521849358 Serial Number: Vital Sync 4.0.1 CFN: VSLBASE04P / VITALSYNCSW04 UDI-DI (GTIN): 10884521856134 Serial Number: Vital Sync 4.0.2

View the original FDA recall notice

Why was this product recalled?

Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.

What actions did Covidien LLC take?

On 01/30/2026, the firm sent via mail or email an "Urgent Medical Device Correction" letter to customers informing them that based on a customer report determined that the SpO2 No Pulse Alarm was not transmitted to the Vital Sync" system as it was not properly configured in the system software. It was further identified that two additional alarms from RespArray" bedside monitors, CO2 No Breath Detected and RESP No Breath Detected Alarms, were not properly configured in the Vital Sync" system software. This issue only affects Vital Sync" remote patient monitoring system used with a RespArray" bedside monitor. Customers are instructed to: "Notify all personnel in all care environments where RespArray" monitors are connected to Vital Sync"systems about this safety notice. "Post this notice in a prominent location and ensure awareness of the issue is maintained until the software update is installed. "If further distribution of RespArray" monitors that are connected to a Vital Sync" remote monitoring system to other facilities, promptly forward a copy of this letter to those recipients. Patient Management: For patients currently monitored by a RespArray" patient monitor and connected to the HealthCast" Vital Sync" remote monitoring system as an adjunct for patient monitoring, be advised that the SpO2 No Pulse, CO2 No Breath Detected and RESP No Breath Detected Alarm notifications from the primary RespArray" patient monitor are not transmitted to the Vital Sync" remote monitoring system. Refer to Clinical Discovery Guide for specific configurations selected by your facility. No additional action is required by treating clinicians other than recognition that the SpO2 No Pulse, CO2 No Breath Detected and RESP No Breath Detected Alarms are not available to be transmitted from the RespArray" monitor to the Vital Sync" remote monitoring system until after the updated software is installed. For question - contact Technical and Clinical Support at 800-255-6774, Option 6.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.