Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1559-2025

Welch Allyn CONNEX Accessory Power Management Stand: Recall

Welch Allyn CONNEX Accessory Power Management Stand:

FDA recall Class II

Ongoing

FDA lists this Class II recall for Welch Allyn CONNEX Accessory Power Management Stand:. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1559-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 24, 2025
FDA report date
Apr 23, 2025
Quantity in commerce
114,202 units
Distribution
See key details
FDA cause
Process design

What is being recalled?

Welch Allyn CONNEX Accessory Power Management Stand:

Affected model & UDI checker

Search the exact identifiers published for Z-1559-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedWelch Allyn CONNEX Accessory Power Management Stand:Check source notice00732094210613
Not specifiedWelch Allyn CONNEX Accessory Power Management Stand:Check source notice00732094240597
Not specifiedWelch Allyn CONNEX Accessory Power Management Stand:Check source notice00732094240603
FDA code information

All Serial Numbers. Product Code: 7000-APM, UDI-DI: 00732094210613; Product Code: 75-HCA-CTB, UDI-DI: 00732094240597; Product Code: 75-HCA-MTB, UDI-DI: 00732094240603;

View the original FDA recall notice

Why was this product recalled?

There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

What actions did Baxter Healthcare Corporation take?

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/24/25 was sent to customers. Actions to be Taken by Customers 1. DO NOT USE UNAPPROVED BATTERIES. Baxter asks customers to only use the approved Baxter / Welch Allyn BATT99 battery, consistent with product labeling due to their safety and performance testing. 2. INSPECT ALL DEVICES. If you have a device identified in the affected product table above, check the device for the correct battery type. See Figure 1 below for examples of labels from approved batteries labelled with the BATT99 product code, Welch Allyn , and the Skaneateles Falls manufacturing facility address. 3. STOP USING THE DEVICE if an unapproved battery is found. 4. IMMEDIATELY REPLACE THE UNAPPROVED BATTERY with an approved battery. Please contact Baxter to order the approved Baxter / Welch Allyn BATT99 batteries. Batteries can be ordered at the following link: https://parts.hillrom.com/hillromUS/en 5. Please forward a copy of this communication to the Medical Director/Office Manager, Director of Biomedical Engineering, Patient Safety Officer, Risk Manager, Surgical Director and any other departments within your institution who use the affected product. 6. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 7. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 8. If you are a dealer/wholesaler, distributor/reseller or Original Equipment Manufac

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology