Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Recall
Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1573-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 18, 2026
- FDA report date
- Mar 25, 2026
- Quantity in commerce
- 59
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package
Affected model & UDI checker
Search the exact identifiers published for Z-1573-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 901029 | Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 | Check source notice | Not specified |
| VS100 | Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 | Check source notice | Not specified |
FDA code information
Lot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table
Why was this product recalled?
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
What actions did Baxter Healthcare Corporation take?
On February 18, 2026, the firm notified consignees via Urgent Medical Device Recall letters. Users were instructed to immediately locate, isolate, and stop using the listed affected products. Customers should contact Baxter Technical Support to arrange for the return and correction of their impacted device. Baxter may provide a loaner device upon request while your device is being upgraded and serviced. Distributors should return unopened products in inventory for credit. If affected product was further distributed, the distributor should conduct a consumer-level recall.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1573-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026