FDA RECALL RECORD · Z-1577-2025

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2). Recall

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

FDA recall Class II

Ongoing

FDA lists this Class II recall for Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1577-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 10, 2025
FDA report date
Apr 23, 2025
Quantity in commerce
110 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

Affected model & UDI checker

Search the exact identifiers published for Z-1577-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
VH-3010Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).Check source noticeNot specified
Not specifiedMaquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).Check source notice0060756770826
FDA code information

Model No. VH-3010; UDI-DI: 0060756770826; Serial No. H24080065G, H24080074G, H24080012G, H24080022G, H24080184G, H24080185G, H24080051G, H24080117G, H24080069G, H24080016G, H24080079G, H24080041G, H24080056G, H24080030G, H24080068G, H24080034G, H24080058G, H24080026G, H24080033G, H24080020G, H24080032G, H24080053G, H24080066G, H24080077G, H24080144G, H24080123G, H24080042G, H24080070G, H24080008G, H24080003G, H24080174G, H24080064G, H24080027G, H24080044G, H24080047G, H24080049G, H24080061G, H24080078G, H24080010G, H24080138G, H24080080G, H24080006G, H24080015G, H24080038G, H24080085G, H24080084G, H24080031G, H24080101G, H24080148G, H24080040G, H24080055G, H24080025G, H24080052G, H24080048G, H24080093G, H24080126G, H24080073G, H24080007G, H23060082G, H24080075G, H24080113G, H24080127G, H24080011G, H24080005G, H24080013G, H24080023G, H24080057G, H24080110G, H24080009G, H24080067G, H24080029G, H24080043G, H24080183G, H24080165G, H24080017G, H24080001G, H24080037G, H24080060G, H24080072G, H24080076G, H24080045G, H24080024G, H24080028G, H24080114G, H24080136G, H24080062G, H24080018G, H24080054G, H24080063G, H24080112G, H24080118G, H24080071G, H24080182G, H24080002G, H24080004G, H24080014G, H24080019G, H24080021G, H24080035G, H24080036G, H24080039G, H24080046G, H24080050G, H24080059G, H24080081G, H24080082G, H24080083G, H24080125G, H24080186G, H24080189G.

View the original FDA recall notice

Why was this product recalled?

Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.

What actions did Maquet Cardiovascular, LLC take?

An URGENT MEDICAL DEVICE REMOVAL notification dated March 2025 was mailed to consignees. The notification instructs consignees to examine their inventory for impacted serial numbers. For power supplies whose serial numbers are not listed in the provided Appendix A of recalled devices, devices can continue to be used. If power supplies with affected serial numbers are identified as being in stock, the associated devices should not be used and returned to Getinge. Returns are facilitated by contacting Getinge Customer Service at 888-800-2874 between the hours of 6:00 AM and 5:00 PM PST to request a return authorization number and shipping instructions. Consignees are to forward the provided notification to all users of the device within the facility and forwarded to any locations where devices were further distributed. Consignees with any questions about the recall are to contact their Getinge representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.