Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1579-2025

Brand Name: Rover Product Name: Mobile X-ray System Recall

Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Rover Product Name: Mobile X-ray System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1579-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 28, 2025
FDA report date
Apr 23, 2025
Quantity in commerce
20 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-1579-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MXU-RV35Brand Name: Rover Product Name: Mobile X-ray SystemCheck source noticeNot specified
FDA code information

Serial Number/UDI: MSN0348 (01)09357123000037(11)230123(21)00348; MSN0349 (01)09357123000037(11)230209(21)00349; MSN0351 (01)09357123000037(11)230413(21)00351; MSN0359 (01)09357123000037(11)231130(21)00359; MSN0361 (01)09357123000037(11)230503(21)00361; MSN0365 (01)09357123000037(11)230510(21)00365; MSN0391 (01)09357123000037(11)230929(21)00391; MSN0406 (01)09357123000037(11)240301(21)00406; MSN0416 (01)09357123000037(11)240429(21)00416; MSN0419 (01)09357123000037(11)240628(21)00419; MSN0428 (01)09357123000037(11)240823(21)00428; MSN0429 (01)09357123000037(11)240823(21)00429; MSN0430 (01)09357123000037(11)240824(21)00430; MSN0431 (01)09357123000037(11)240826(21)00431; MSN0432 (01)09357123000037(11)241108(21)00432; MSN0434 (01)09357123000037(11)241115(21)00434; MSN0435 (01)09357123000037(11)241118(21)00435; MSN0436 (01)09357123000037(11)241122(21)00436; MSN0437 (01)09357123000037(11)241125(21)00437; MSN0438 (01)09357123000037(11)250114(21)00438

View the original FDA recall notice

Why was this product recalled?

Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Micro-X Ltd. take?

On March 28, 2025, Micro-X Ltd. issued a "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. Micro-X asked consignees to take the following actions: 1. Continue to use your product as normal in accordance with its intended use. 2. Your local service representative will be in contact to apply this update to the systems at the next site visit. 3. Circulate this notice to all users of this device so they are aware of the issue. 4. If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. 5. Please retain this letter with your system(s) documentation. 6. Please complete and return the attached response form to Micro-X to the Micro-X LTD in Australia: support@micro-x.com and copy djerkovic@micro-x.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history