Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Recall
Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1586-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 18, 2026
- FDA report date
- Mar 25, 2026
- Quantity in commerce
- 306
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-1586-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 303585 | Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay | Check source notice | Not specified |
FDA code information
Lot Code: The Aptima HPV Assay Kit Part Number 303585, Master Lot 929785 contains one Refrigerated Box (PN 302883) and one Room Temperature Box (PN 302885). The following are the UDI number for each box: UDI Number for PN 302883: 15420045500051 UDI Number for PN 302885: 15420045500068
Why was this product recalled?
Due to product exhibiting potential to generate either invalid or false negative results.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Hologic, Inc. take?
On 02/18/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter to customers informing them that Hologic has identified that the Aptima HPV Assay kits from ML 929785 exhibit the potential to generate either invalid results for samples and calibrators or false negative results for low analyte samples when reagents are not used on the day of reconstitution. Customers are instructed: 1. Discontinue use of the Aptima HPV Assay kits from ML 929785. 2. Segregate and count the number of affected product on-hand and dispose according t the local guidelines and regulations. 3.Fill out the Customer Response Form using the link https://iqvia-response.my.site.com/mt/fca?cid=hpv-929785 4. Hologic will process and ship the replacement kits For Questions or concerns, contact Hologic Technical Support at +1 888-484-4747 or +1 858-410-8511 during hours 5 am to 5 pm PT or by email at molecularsupport@hologic.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1586-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology