Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12. Recall
Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1597-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 23, 2026
- FDA report date
- Apr 1, 2026
- Quantity in commerce
- 70 units
- Distribution
- See key details
- FDA cause
- No Marketing Application
What is being recalled?
Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
1 Article number (Lot Numbers): 21112-7 (643/0549), 21114-5 (643/032301, 643/042303, 643/042304, 643/0618, 643/102211, 643/112304), 21113-9 (643/072320). 2. Article number (Lot Numbers): 21111-6 (643/032306, 643/072319, 643/112201), 21111-8 (643/032318), 21111-12 (643/032305).
Why was this product recalled?
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
What actions did I.T.S. GmbH take?
ITS initially notified its importer/distributor on about 01/23/2026 via email. Instructions were to cease distribution, and further information will be provided. An additional notification was sent to distributors and consignees on about 02/04/2026 via email. Instructions included to ensure any IFU different to the provided one is removed from systems and/or destroyed, notify clinical staff managing patients with implants, contact I.T.S. GmbH for any questions or to discuss proper scanning conditions for implanted devices in the interim while they work on labeling and IFU updates for the devices, and arrange for the return of specific originally packaged items - base plates and screws in surgery trays. Other outlined products can continue to be used upon confirmation from ITS. Consignees were also instructed to complete and return the response form. Distributors were instructed to continue to immediately halt product distribution from the U.S. office to sales groups and hospitals, remove the outdated paper IFU from packaging and add the eIFU symbol to product labeling to implement the updated IFU provided electronically - see annex "F-7.5.5-3-US_IFU_non-sterile_Rev00_DRAFT". Paper IFUs will be made available upon request. They were also requested to assist in notification of hospitals.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1597-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026