MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli. Recall
MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
FDA recall Class II
Ongoing
FDA lists this Class II recall for MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1609-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 24, 2025
- FDA report date
- Apr 1, 2026
- Quantity in commerce
- 685 boxes
- Distribution
- See key details
- FDA cause
- Material/Component Contamination
What is being recalled?
MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Affected model & UDI checker
Search the exact identifiers published for Z-1609-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| C42464 | MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli. | Check source notice | Not specified |
| Not specified | MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli. | Check source notice | 15099590731212 |
FDA code information
Catalog Number: C42464 UDI-DI code: 15099590731212 Lot Number: 2026-01-03
Why was this product recalled?
Due to the likely presence of contamination in well(s).
What actions did BECKMAN COULTER, INC. take?
On 07/24/2025, the firm sent via mail/email an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter has received reports that when testing clinical isolates there was no growth observed in Ceftazidime (Caz) 1, 8 and 16 mg/mL wells; however, growth was observed in the 4 mg/mL well with multiple panels from this lot. Upon further investigation, it was confirmed the growth was Bacillus spp. and not the clinical isolates being tested. Customers are instructed to: -Consider reviewing previous Caz antimicrobial test results from this lot for any spot growth detection (e.g., isolates with no growth in in Caz 1, 8 and 16 mg/mL wells and growth in the 4 mg/mL well or if there is an unusual increase in Caz MIC values). -Discontinue use of MicroScan Neg MIC 56 Lot 2026-01-03 For any questions regarding this notice, contact Customer Technical Support: -From our website: http://www.beckmancoulter.com -By phone: call 800 677-7226 option 1 in the United States. If replacement product is needed: -Complete the attached Replacement Order Form and email to askbeckman@beckman.com or fax to (866) 294-7850 OR -Call Client Services at (800) 526-3821
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1609-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026