Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1641-2025

Merit Medical , REF: R2275, EsophyX Z+ with SerosaFuse Implantable Fasteners, (1) Fastener Delivery Device, (2) 7.5mm Cartridges, STERILEEO,RxOnly Recall

Merit Medical , REF: R2275, EsophyX Z+ with SerosaFuse Implantable Fasteners, (1) Fastener Delivery Device, (2) 7.5mm Cartridges, STERILEEO,RxOnly

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merit Medical , REF: R2275, EsophyX Z+ with SerosaFuse Implantable Fasteners, (1) Fastener Delivery Device, (2) 7.5mm Cartridges, STERILEEO,RxOnly. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1641-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 14, 2025
FDA report date
Apr 30, 2025
Quantity in commerce
7453 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Merit Medical , REF: R2275, EsophyX Z+ with SerosaFuse Implantable Fasteners, (1) Fastener Delivery Device, (2) 7.5mm Cartridges, STERILEEO,RxOnly

Affected model & UDI checker

Search the exact identifiers published for Z-1641-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
R2275Merit Medical , REF: R2275, EsophyX Z+ with SerosaFuse Implantable Fasteners, (1) Fastener Delivery Device, (2) 7.5mm Cartridges, STERILEEO,RxOnlyCheck source noticeNot specified
Not specifiedMerit Medical , REF: R2275, EsophyX Z+ with SerosaFuse Implantable Fasteners, (1) Fastener Delivery Device, (2) 7.5mm Cartridges, STERILEEO,RxOnlyCheck source notice00810275014011
FDA code information

All Lots/ UDI: 00810275014011

View the original FDA recall notice

Why was this product recalled?

Update to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners.

What actions did Merit Medical Systems, Inc. take?

On March 14, 2025, Merit Medical issued a "Urgent Medical Device Recall Notice via E-Mail, followed by mail to affected consignees. Merit asked consignees to take the following actions: 1. Read and understand this communication and the updated IFU referenced in the attached poster. 2. There is no need to return product to Merit. 3. Post the attached poster on or near the affected products so users are aware of the updated IFU. 4. Discard the current IFU with the product at point of use. 5. Ensure that applicable personnel within your organization are made aware of this communication. 6. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 7. If you require paper IFUs, please contact RESPONSE@merit.com. 8. Please acknowledging you are aware of the information communicated and have completed the actions required of you. 9. If you require additional product training regarding the IFU updates, please contact your Merit Sales Representative to schedule.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.