Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Recall

Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1642-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 25, 2026
FDA report date
Apr 1, 2026
Quantity in commerce
93 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Affected model & UDI checker

Search the exact identifiers published for Z-1642-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
3005PKBrand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting ForcepsCheck source notice00821925036000
FDA code information

Model: 3005PK; UDI: 00821925036000; Lot Numbers: All unexpired;

View the original FDA recall notice

Why was this product recalled?

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

What actions did Olympus Corporation of the Americas take?

On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this letter through the Olymps portal 5. Please forward this notice to other users who may have the affected products if you have further distributed it.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology