InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA Recall
InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
FDA recall Class II
Ongoing
FDA lists this Class II recall for InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1650-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 4, 2025
- FDA report date
- Apr 30, 2025
- Quantity in commerce
- 8,986
- Distribution
- See key details
- FDA cause
- Mixed-up of materials/components
What is being recalled?
InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
Affected model & UDI checker
Search the exact identifiers published for Z-1650-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MMT-105NNBLNA | InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA | Check source notice | Not specified |
| MMT-105NNGYNA | InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA | Check source notice | Not specified |
| MMT-105NNPKNA | InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA | Check source notice | Not specified |
| MMT-105ELBLNA | InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA | Check source notice | Not specified |
| MMT-105ELGYNA | InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA | Check source notice | Not specified |
| MMT105ELPKNA | InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA | Check source notice | Not specified |
FDA code information
REF/UDI-DI(GTIN)/Lot: MMT-105NNBLNA/763000827182/D0020, D0034, D0036, D0038, D0045, D0046, D0050, D0057; MMT-105NNGYNA/763000827199/D0026, D0037, D0039, D0041, D0042, D0043, D0049; MMT-105NNPKNA/763000827205/D0029, D0030, D0040, D0051, D0062; MMT-105ELBLNA/763000827151/D0021, D0033, D0035, D0044, D0059, D0060, D0061; MMT-105ELGYNA/763000827168/D0031, D0032, D0047, D0068; MMT105ELPKNA/763000827175/D0027, D0028, D0048
Why was this product recalled?
Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove from the insulin pen. If the issue is present, and if the user does not have backup insulin they may experience temporary hyperglycemia.
What actions did Medtronic MiniMed, Inc. take?
On 3/4/2025, correction notices were sent to customers and distributors informing them of the following: 1) If you have already inserted your insulin cartridge into your device without difficulty removing the cartridge holder, then you can continue normal use. 2) If you have a device from one of the affected lot numbers and have not yet started using it, or have had difficulties in removing the cartridge holder or inserting the insulin cartridge, please call 1-800-646-4633, option 1, or go to https://info.medtronicdiabetes.com/ICH-replacement-form to request a replacement product. Do not use this device and dispose of it according to local regulations. 3) Complete and return the confirmation form via link: https://info.medtronicdiabetes.com/cartridge-holder Distributors are asked to notify impacted customers. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1650-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology