RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00 Recall
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
FDA recall Class II
Ongoing
FDA lists this Class II recall for RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1656-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 9, 2026
- FDA report date
- Apr 8, 2026
- Quantity in commerce
- 12
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Affected model & UDI checker
Search the exact identifiers published for Z-1656-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| RXM1000 | RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00 | Check source notice | Not specified |
| Not specified | RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00 | Check source notice | 00860003983812 |
FDA code information
Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.
Why was this product recalled?
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Reflexion Medical, Inc. take?
On 2/9/2026, recall notices were emailed to customers who were informed of the following: How to recognize that the device may fail: Any patient treatment with a roll offset applied in a localization followed by a subsequent repeat localization within the same treatment session will be affected. For patient treatments that have already occurred, measuring the approximate distance off-axis (R) and the maximum roll offset value (., in radians) can provide an approximate displacement value (S) through a simple mathematical formula, which may be helpful in estimating the severity of the dose impact for the patient. Customers were asked to do the following: 1) Notify your clinical staff of the recall, including posting this notification on or near the system until this issue can be resolved. 2) Halt utilization of the Repeat Localization feature for all treatment and imaging sessions. a. Note: Standard Roll corrections may still be used on the system if they are only applied in the localization immediately prior to the treatment delivery or imaging session commencing, as those are unaffected by this issue. 3) If a Roll correction is applied in an initial Localization and a Repeat Localization is initiated, the user should end the treatment session without continuing to treatment delivery or PET imaging. The user should restart the treatment session as a new treatment session to properly apply the Roll correction in the initial Localization without performing a Repeat Localization afterwards. 4) Complete and return the acknowledgement form. Questions or additional assistance can be sent to the firm at support@reflexion.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1656-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology