Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1664-2025

Beckman Coulter DxI 9000 Access Immunoassay Analyzer Recall

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

FDA recall Class II

Ongoing

FDA lists this Class II recall for Beckman Coulter DxI 9000 Access Immunoassay Analyzer. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1664-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 27, 2025
FDA report date
Apr 30, 2025
Quantity in commerce
370 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Affected model & UDI checker

Search the exact identifiers published for Z-1664-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedBeckman Coulter DxI 9000 Access Immunoassay AnalyzerCheck source notice15099590732103
FDA code information

UDI/DI 15099590732103, all units run with system software versions SW 1.20.0 and below

View the original FDA recall notice

Why was this product recalled?

When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.

What actions did BECKMAN COULTER, INC. take?

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/27/2025 via USPS first class mail. The notice explained the issue with the device, impact to the patient, and provided a workaround until a software upgrade is available. If the consignee further distributed the affected device, they were directed to provide a copy of the notice to the recipient. For questions, contact Customer Support Center, from the website: http://www.beckmancoulter.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology