epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n. Recall
epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
FDA recall Class II
Ongoing
FDA lists this Class II recall for epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1685-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 19, 2025
- FDA report date
- May 7, 2025
- Quantity in commerce
- 5,207 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
Affected model & UDI checker
Search the exact identifiers published for Z-1685-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n. | Check source notice | 00809708072254 |
FDA code information
Siemens Material Number (SMN): 10736382; Udi-DI: 00809708072254; Lot Numbers: All lot numbers with prefix (xx) 00 or 06; epoc Software Version and Senor Configuration: epoc Host SW V3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
Why was this product recalled?
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
What actions did Siemens Healthcare Diagnostics, Inc. take?
Consignees were sent an Urgent Medical Device Correction notice dated March 2025. The notification instructs consignees to review the provided notice with their medical director to determine the appropriate course of action. If affected devices were further distributed consignees are asked to forward the recall notification. Consignees are to return the provided Field Correction Effectiveness Check form within 7 days of receipt. The firm recommends that consignees stop injecting the sample as soon as the audio/visual cues are presented to minimize the potential for erroneous pH results. The current affected software expires on 06/10/25 and the upgrade will include a correction for this issue. Consignees are to retain the recall notification with their manual or near the system for reference. Consignees with questions can call the Siemens Healthineers Technical Solutions Center at 877-229-3711.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1685-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology