Distal Femoral Augment; Item Numbers: (1) 880-300/11 (Sz. 0 - cemented sm), (2) 880-300/21 (Sz. 0 - cemented lg), (3) 880-301/11 (Sz. 1-2 - cemented sm), (4) 880-301/21 (Sz. 1-2 - cemented lg), (5) 880-303/11 (Sz. 3-4 - cemented sm), (6) 880-303/12 (Sz. 3-4 - cemented md), (7) 880-303/21 (Sz. 3-4 - cemented lg), (8) 880-303/22 (Sz. 3-4 - cemented x-lg), (9) 880-305/11 (Sz. 5-6 - cemented sm), (10) 880-305/12 (Sz. 5-6 - cemented md), (11) 880-305/21 (Sz. 5-6 - cemented lg), (12) 880-305/22 (Sz. 5-6 - cemented x-lg), (13) 880-307/11 (Sz. 7-8 - cemented sm), (14) 880-307/12 (Sz. 7-8 - cemented md), (15) 880-307/21 (Sz. 7-8 - cemented lg), (16) 880-307/22 (Sz. 7-8 - cemented x-lg), (17) 880-309/11 (Sz. 9-10 - cemented sm), (18) 880-309/12 (Sz. 9-10 - cemented md), (19) 880-309/21 (Sz. 9-10 - cemented lg), (20) 880-309/22 (Sz. 9-10 - cemented x-lg). Recall
Distal Femoral Augment; Item Numbers: (1) 880-300/11 (Sz. 0 - cemented sm), (2) 880-300/21 (Sz. 0 - cemented lg), (3) 880-301/11 (Sz. 1-2 - cemented sm), (4) 880-301/21 (Sz. 1-2 - cemented lg), (5) 880-303/11 (Sz. 3-4 - cemented sm), (6) 880-303/12 (Sz. 3-4 - cemented md), (7) 880-303/21 (Sz. 3-4 - cemented lg), (8) 880-303/22 (Sz. 3-4 - cemented x-lg), (9) 880-305/11 (Sz. 5-6 - cemented sm), (10) 880-305/12 (Sz. 5-6 - cemented md), (11) 880-305/21 (Sz. 5-6 - cemented lg), (12) 880-305/22 (Sz. 5-6 - cemented x-lg), (13) 880-307/11 (Sz. 7-8 - cemented sm), (14) 880-307/12 (Sz. 7-8 - cemented md), (15) 880-307/21 (Sz. 7-8 - cemented lg), (16) 880-307/22 (Sz. 7-8 - cemented x-lg), (17) 880-309/11 (Sz. 9-10 - cemented sm), (18) 880-309/12 (Sz. 9-10 - cemented md), (19) 880-309/21 (Sz. 9-10 - cemented lg), (20) 880-309/22 (Sz. 9-10 - cemented x-lg).
FDA recall Class II
Ongoing
FDA lists this Class II recall for Distal Femoral Augment; Item Numbers: (1) 880-300/11 (Sz. 0 - cemented sm), (2) 880-300/21 (Sz. 0 - cemented lg), (3) 880-301/11 (Sz. 1-2 - cemented sm), (4) 880-301/21 (Sz. 1-2 - cemented lg), (5) 880-303/11 (Sz. 3-4 - cemented sm), (6) 880-303/12 (Sz. 3-4 - cemented md), (7) 880-303/21 (Sz. 3-4 - cemented lg), (8) 880-303/22 (Sz. 3-4 - cemented x-lg), (9) 880-305/11 (Sz. 5-6 - cemented sm), (10) 880-305/12 (Sz. 5-6 - cemented md), (11) 880-305/21 (Sz. 5-6 - cemented lg), (12) 880-305/22 (Sz. 5-6 - cemented x-lg), (13) 880-307/11 (Sz. 7-8 - cemented sm), (14) 880-307/12 (Sz. 7-8 - cemented md), (15) 880-307/21 (Sz. 7-8 - cemented lg), (16) 880-307/22 (Sz. 7-8 - cemented x-lg), (17) 880-309/11 (Sz. 9-10 - cemented sm), (18) 880-309/12 (Sz. 9-10 - cemented md), (19) 880-309/21 (Sz. 9-10 - cemented lg), (20) 880-309/22 (Sz. 9-10 - cemented x-lg).. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1699-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 3, 2025
- FDA report date
- May 7, 2025
- Quantity in commerce
- 1180 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Distal Femoral Augment; Item Numbers: (1) 880-300/11 (Sz. 0 - cemented sm), (2) 880-300/21 (Sz. 0 - cemented lg), (3) 880-301/11 (Sz. 1-2 - cemented sm), (4) 880-301/21 (Sz. 1-2 - cemented lg), (5) 880-303/11 (Sz. 3-4 - cemented sm), (6) 880-303/12 (Sz. 3-4 - cemented md), (7) 880-303/21 (Sz. 3-4 - cemented lg), (8) 880-303/22 (Sz. 3-4 - cemented x-lg), (9) 880-305/11 (Sz. 5-6 - cemented sm), (10) 880-305/12 (Sz. 5-6 - cemented md), (11) 880-305/21 (Sz. 5-6 - cemented lg), (12) 880-305/22 (Sz. 5-6 - cemented x-lg), (13) 880-307/11 (Sz. 7-8 - cemented sm), (14) 880-307/12 (Sz. 7-8 - cemented md), (15) 880-307/21 (Sz. 7-8 - cemented lg), (16) 880-307/22 (Sz. 7-8 - cemented x-lg), (17) 880-309/11 (Sz. 9-10 - cemented sm), (18) 880-309/12 (Sz. 9-10 - cemented md), (19) 880-309/21 (Sz. 9-10 - cemented lg), (20) 880-309/22 (Sz. 9-10 - cemented x-lg).
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Item Number/UDI-DI: (1) 880-300/11 (04026575284931), (2) 880-300/21 (04026575257102), (3) 880-301/11 (04026575257133), (4) 880-301/21 (04026575257140), (5) 880-303/11 (04026575257171), (6) 880-303/12 (04026575257188), (7) 880-303/21 (04026575257195), (8) 880-303/22 (04026575257201), (9) 880-305/11 (04026575257256), (10) 880-305/12 (04026575257263), (11) 880-305/21 (04026575257270), (12) 880-305/22 (04026575257287), (13) 880-307/11 (04026575257379), (14) 880-307/12 (04026575257386), (15) 880-307/21 (04026575257393), (16) 880-307/22 (04026575257409), (17) 880-309/11 (04026575257454), (18) 880-309/12 (04026575257461), (19) 880-309/21 (04026575257478), (20) 880-309/22 (04026575257485). Applies to all lots of the LINK SymphoKnee Femoral Augments listed, manufactured up until 03/01/2025.
Why was this product recalled?
The device was delivered with a preassembled Femoral Augment screw that was missing its thread.
What actions did Waldemar Link GmbH & Co. KG (Mfg Site) take?
On April 3, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were emailed to customers. Actions to be taken by the Customer/User: Please take note of the following instructions for replacing any screws found missing the thread. No product return is required. Instructions for disassembling and replacing a defective screw: Disassembly 1. Use the 2.5 mm Torque Wrench article REF (15-2545) to unscrew the screw from the Augment while pushing the tip of the screw towards the Augment. 2. Exchange the screw. Open an alternative Augment and use its screw. Alternative sizes of Augments are always available in the operating room during the surgery. For all Femoral Augment sizes and types the same long screw type is preassembled and for all Tibial Augments the same short screw type is preassembled, see table 1. Assembly/Replacement 3. Use the 2.5 mm Torque Wrench article REF (15-2545) to screw in the replacement screw into the Augment. 4. The screw head must be countersunk in the Augment and the screw tip must protrude. Follow the LinkSymphoKnee surgical technique for all other steps. Please respond using the Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the instructions provided above to replace any affected screws. Type of Action by the Company: LINK will provide sterilized replacement screws (for both Femoral Augments and Tibial Augments) for each affected distributor as a back-up for upcoming surgeries. Other Information: Opening an additional Augment to exchange the screw will not incur any costs to you. Should you have any questions about acquiring backup/replacement screws for forthcoming surgeries (once available), please contact Customer Service at LinkBio (order@linkbio.com). Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any additional questions.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1699-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology