Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1703-2025

myQA iON; Article Number: MQ10-000; Recall

myQA iON; Article Number: MQ10-000;

FDA recall Class II

Ongoing

FDA lists this Class II recall for myQA iON; Article Number: MQ10-000;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1703-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 1, 2025
FDA report date
May 7, 2025
Quantity in commerce
84 units (9US, 75 OUS)
Distribution
See key details
FDA cause
Software design

What is being recalled?

myQA iON; Article Number: MQ10-000;

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Article Number: MQ10-000; UDI-DI: EIBAMQ010000;

View the original FDA recall notice

Why was this product recalled?

Wrong analysis results can occur if the user creates two structures with the same name, except for one differing by a special character.

What actions did IBA Dosimetry GmbH take?

On April 1, 2025, FIELD SAFETY NOTICE letters were sent to customers. Actions to be taken to prevent the error from occurring: When naming structures, please refrain from using any of the following special characters: % \ / : * ? < > | + Additionally, during the export of files into myQA iON, some TPSs cut off the structure names due to their length. Users are not notified when names are automatically shortened by the system. Please check that the structure name was correctly imported. Distribution of this Field Safety Notice: Please distribute this notice to all those who need to be aware within your area of responsibility where the affected medical device is in use. We would kindly ask you to acknowledge the receipt of this Field Safety Notice. Please sign and send back the attached Confirmation of Receipt within four weeks after the receipt of this Field Safety Notice. Contact information: Please feel free to contact us as follows if you have any further questions: USA, Canada, Latin America Europe, Middle East, Africa Asia Pacific Phone: +1 786 288 0369 Phone: +49 9128 607 38 Phone: +65 3129 2472 service-usa@iba-group.com service-emea@iba-group.com service-apac@iba-group.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology