Meridian Bioscience ImmunoCard STAT! Crypto/Giardia Test Kits, Model Number 750830 Recall
Meridian Bioscience ImmunoCard STAT! Crypto/Giardia Test Kits, Model Number 750830
FDA recall Class II
Ongoing
FDA lists this Class II recall for Meridian Bioscience ImmunoCard STAT! Crypto/Giardia Test Kits, Model Number 750830. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1719-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 24, 2025
- FDA report date
- May 14, 2025
- Quantity in commerce
- 56 boxes
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Meridian Bioscience ImmunoCard STAT! Crypto/Giardia Test Kits, Model Number 750830
Affected model & UDI checker
Search the exact identifiers published for Z-1719-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 750830 | Meridian Bioscience ImmunoCard STAT! Crypto/Giardia Test Kits, Model Number 750830 | Check source notice | Not specified |
| Not specified | Meridian Bioscience ImmunoCard STAT! Crypto/Giardia Test Kits, Model Number 750830 | Check source notice | 00742860100017 |
FDA code information
UDI-DI: 00742860100017; Lot Number: 08241275
Why was this product recalled?
A single lot of the ImmunoCard STAT! Crypto/Giardia Test Kit was inadvertently stored in Medline warehouses at temperatures outside of the labeled storage requirements.
What actions did MEDLINE INDUSTRIES, LP - Northfield take?
An URGENT MEDICAL DEVICE RECALL notification letter dated 3/24/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Please destroy any affected inventory. Medline will issue credit for the product based upon the quantity and lot numbers identified on the submitted response form The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-25-063 Recall Code: Recall Code 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers return any affected product to you for collection and return to Medline Industries, LP. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1719-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology