FDA RECALL RECORD · Z-1721-2025

Washer Disinfector Aquadis 56; Recall

Washer Disinfector Aquadis 56; Model Number: 56M

FDA recall Class II

Ongoing

FDA lists this Class II recall for Washer Disinfector Aquadis 56;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1721-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 2, 2025
FDA report date
May 14, 2025
Quantity in commerce
424 units (79 US, 345 OUS)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Washer Disinfector Aquadis 56; Model Number: 56M

Affected model & UDI checker

Search the exact identifiers published for Z-1721-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
56MWasher Disinfector Aquadis 56;Check source noticeNot specified
Not specifiedWasher Disinfector Aquadis 56;Check source notice07340153710351
FDA code information

Model: 56M; UDI-DI: 07340153710351; Serial Numbers: WAA105199, WAA105251, WAA105268, WAA105357, WAA105535, WAA105700, WAA105915, WAA105972, WAA105973, WAA106079, WAA106114, WAA106135, WAA106451, WAA106471, WAA106487, WAA106505, WAA107436, WAA107492, WAA107565, WAA107581, WAA107588, WAA107920, WAA108225, WAA108807, WAA108920, WAA108921, WAA108922, WAA109036, WAA109047, WAA109231, WAA109377, WAA109392, WAA109635, WAA109687, WAA109731, WAA109790, WAA109835, WAA109840, WAA109853, WAA109907, WAA109972, WAA110109, WAA110366, WAA110487, WAA110546, WAA110557, WAA110711, WAA110726, WAA110802, WAA110847, WAA110862, WAA110984, WAA110999, WAA111007, WAA111011, WAA111068, WAA111086, WAA111087, WAA111200, WAA600033, WAA600054, WAA600100, WAA600102, WAA600103, WAA600145, WAA600190, WAA600282, WAA600369, WAA600431, WAA600434, WAA600444, WAA600445, WAA600553, WAA600554, WAA600587, WAA600588, WAA600609, WAA600610, WAA600704, WAA106987, WAA110250, WAA110257, WAA108634, WAA110823, WAA108206, WAA109799, WAA109066, WAA109076, WAA109091, WAA109373, WAA110528, WAA111028, WAA600134, WAA600189, WAA110820, WAA110123, WAA108871, WAA108872, WAA106724, WAA109403, WAA109405, WAA109495, WAA109933, WAA110513, WAA110527, WAA110617, WAA109658, WAA109672, WAA110029, WAA110779, WAA110893, WAA107614, WAA108608, WAA108610, WAA109222, WAA109764, WAA109782, WAA109946, WAA110055, WAA110083, WAA110204, WAA110801, WAA600066, WAA600068, WAA600072, WAA600507, WAA600508, WAA106520, WAA106618, WAA106639, WAA110594, WAA110595, WAA110842, WAA111150, WAA600105, WAA108702, WAA108040, WAA106692, WAA107486, WAA107491, WAA108832, WAA108998, WAA109241, WAA109244, WAA109885, WAA110167, WAA110279, WAA110290, WAA110303, WAA110349, WAA110394, WAA110406, WAA111093, WAA111098, WAA111112, WAA600216, WAA600441, WAA600509, WAA600560, WAA600611, WAA600612, WAA600613, WAA600648, WAA107279, WAA107940, WAA109140, WAA109973, WAA110534, WAA110545, WAA111142, WAA111174, WAA111177, WAA111185, WAA106240, WAA106760, WAA110566, WAA110876, WAA108501, WAA108502, WAA108514, WAA109093, WAA109409, WAA109805, WAA109808, WAA109966, WAA110032, WAA110037, WAA110618, WAA600141, WAA600220, WAA600221, WAA600222, WAA600223, WAA600253, WAA600254, WAA600255, WAA600450, WAA600451, WAA600585, WAA600586, WAA110881, WAA105641, WAA105682, WAA106980, WAA107125, WAA108151, WAA108367, WAA108394, WAA108519, WAA108531, WAA108953, WAA109653, WAA109874, WAA110018, WAA110027, WAA110056, WAA110327, WAA111114, WAA600106, WAA600186, WAA600187, WAA107148, WAA107232, WAA107267, WAA107307, WAA107360, WAA107418, WAA108067, WAA108136, WAA108137, WAA108161, WAA108170, WAA108217, WAA108409, WAA108479, WAA108499, WAA108500, WAA108653, WAA108743, WAA108745, WAA108746, WAA108863, WAA108864, WAA109187, WAA109199, WAA109200, WAA109207, WAA109269, WAA109278, WAA109285, WAA109305, WAA109309, WAA109314, WAA109327, WAA109497, WAA109501, WAA109507, WAA109565, WAA110208, WAA110226, WAA110361, WAA110643, WAA110836, WAA110845, WAA110937, WAA110938, WAA111012, WAA111032, WAA111055, WAA600047, WAA600053, WAA600135, WAA600256, WAA600286, WAA600287, WAA600288, WAA600370, WAA600371, WAA600436, WAA600438, WAA600442, WAA600446, WAA600448, WAA600555, WAA110297, WAA110315, WAA107182, WAA109948, WAA110950, WAA110268, WAA110648, WAA110663, WAA107190, WAA107199, WAA109370, WAA109513, WAA110486, WAA106247, WAA106399, WAA106412, WAA108983, WAA110013, WAA110420, WAA600217, WAA600649, WAA600858, WAA109765, WAA109696, WAA109713, WAA109737, WAA109182, WAA109673, WAA109686, WAA109862, WAA109864, WAA110256, WAA110274, WAA110460, WAA109949, WAA106546, WAA106548, WAA107084, WAA107905, WAA108894, WAA109343, WAA109352, WAA109393, WAA109908, WAA109931, WAA110686, WAA110833, WAA111083, WAA600218, WAA600219, WAA600278, WAA600556, WAA107522, WAA107594, WAA108973, WAA108975, WAA110638, WAA111052, WAA111053, WAA111186, WAA111189, WAA110395, WAA107136, WAA110323, WAA110338, WAA110360, WAA110370, WAA107215, WAA106992, WAA108059, WAA108623, WAA110703, WAA105696, WAA107624, WAA107641, WAA107691, WAA107699, WAA107731, WAA107736, WAA107746, WAA107756, WAA107763, WAA107788, WAA107790, WAA107801, WAA107804, WAA107806, WAA107825, WAA107829, WAA107858, WAA107860, WAA107883, WAA107897, WAA108226, WAA108264, WAA108267, WAA108300, WAA108301, WAA108410, WAA108466, WAA108952, WAA109249, WAA109260, WAA109261, WAA109266, WAA109340, WAA109688, WAA109757, WAA109937, WAA110374, WAA110428, WAA110432, WAA110452, WAA110453, WAA600011, WAA600082, WAA600101, WAA600104, WAA600107, WAA600129, WAA600130, WAA600285, WAA600551, WAA600552, WAA600559, WAA109920, WAA109929, WAA600280, WAA600281, WAA110756, WAA110757, WAA109631, WAA109916, WAA110235, WAA110248, WAA600289, WAA600327, WAA110154, WAA600650, WAA107607.

View the original FDA recall notice

Why was this product recalled?

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

What actions did Getinge Disinfection Ab take?

On April 2, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions 1. Check the serial number of yourWasher Disinfector Aquadis 56 to determine if it is affected. The serial number can be found on the marking label, located on the inside of the front Panel door. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 3. Complete and return the enclosed response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. Note: If your unit is covered by a service agreement, Getinge will carry out the testing and required rework, if necessary, during the next regularly scheduled maintenance. If your facility currently does not have a service agreement with Getinge, our field service technicians will contact you to set up testing. Getinge will conduct testing to determine if the frequency inverter is programmed or not, and if it is found to be affected by this issue, the memory module in the frequency inverter will be replaced, free of charge. Actions to be taken by Getinge: Getinge will conduct testing to determine if the frequency inverter is programmed or not, and if it is found to be affected by this issue, the memory module in the frequency inverter will be replaced, free of charge. This will be performed during your routine planned maintenance to avoid any inconvenience. Additional Information: Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax using the following: - Online: www.accessdata.fda.gov/scripts/medwatch/ - Regular Mail: Download form at www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form - Fax: 1-800-FDA-0178 (1-800-332-0178) We deeply regret this inconve

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology