Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1722-2025

Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity , Adt/Pedi Intub CO2 Line LTerm Lng , Adt/Pedi Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line High Humidity , Neo/Inf Intub CO2 Line LTerm Lng , VitaLine H Set Adult/Pediatric , VitaLine H Set Infant/Neonatal , FilterLine Set Long Adult/Pediatric , FilterLine H Set Long Adult/Pediatric , FilterLine H Set Long Infant/Neonatal , Trade Compliant: FilterLine H, Adult/Ped , Trade Compliant: FilterLine H, Infant/Neo , FilterLine Set Adult/Pedi , FilterLine H Set Adult/Pedi and FilterLine H Set Infant/Neonatal. Carbon dioxide gas analyzer. Recall

Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity , Adt/Pedi Intub CO2 Line LTerm Lng , Adt/Pedi Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line High Humidity , Neo/Inf Intub CO2 Line LTerm Lng , VitaLine H Set Adult/Pediatric , VitaLine H Set Infant/Neonatal , FilterLine Set Long Adult/Pediatric , FilterLine H Set Long Adult/Pediatric , FilterLine H Set Long Infant/Neonatal , Trade Compliant: FilterLine H, Adult/Ped , Trade Compliant: FilterLine H, Infant/Neo , FilterLine Set Adult/Pedi , FilterLine H Set Adult/Pedi and FilterLine H Set Infant/Neonatal. Carbon dioxide gas analyzer.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity , Adt/Pedi Intub CO2 Line LTerm Lng , Adt/Pedi Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line High Humidity , Neo/Inf Intub CO2 Line LTerm Lng , VitaLine H Set Adult/Pediatric , VitaLine H Set Infant/Neonatal , FilterLine Set Long Adult/Pediatric , FilterLine H Set Long Adult/Pediatric , FilterLine H Set Long Infant/Neonatal , Trade Compliant: FilterLine H, Adult/Ped , Trade Compliant: FilterLine H, Infant/Neo , FilterLine Set Adult/Pedi , FilterLine H Set Adult/Pedi and FilterLine H Set Infant/Neonatal. Carbon dioxide gas analyzer.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1722-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 7, 2025
FDA report date
May 14, 2025
Quantity in commerce
486,631 units
Distribution
See key details
FDA cause
Unknown/Undetermined by firm

What is being recalled?

Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity , Adt/Pedi Intub CO2 Line LTerm Lng , Adt/Pedi Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line High Humidity , Neo/Inf Intub CO2 Line LTerm Lng , VitaLine H Set Adult/Pediatric , VitaLine H Set Infant/Neonatal , FilterLine Set Long Adult/Pediatric , FilterLine H Set Long Adult/Pediatric , FilterLine H Set Long Infant/Neonatal , Trade Compliant: FilterLine H, Adult/Ped , Trade Compliant: FilterLine H, Infant/Neo , FilterLine Set Adult/Pedi , FilterLine H Set Adult/Pedi and FilterLine H Set Infant/Neonatal. Carbon dioxide gas analyzer.

Affected model & UDI checker

Search the exact identifiers published for Z-1722-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
989803204511Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity , Adt/Pedi Intub CO2 Line LTerm Lng , Adt/Pedi Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line High Humidity , Neo/Inf Intub CO2 Line LTerm Lng , VitaLine H Set Adult/Pediatric , VitaLine H Set Infant/Neonatal , FilterLine Set Long Adult/Pediatric , FilterLine H Set Long Adult/Pediatric , FilterLine H Set Long Infant/Neonatal , Trade Compliant: FilterLine H, Adult/Ped , Trade Compliant: FilterLine H, Infant/Neo , FilterLine Set Adult/Pedi , FilterLine H Set Adult/Pedi and FilterLine H Set Infant/Neonatal. Carbon dioxide gas analyzer.Check source noticeNot specified
FDA code information

Model No. 989803204511, 989803204321, 989803204301, 989803204331, 989803204521, 989803204531, 989803204311, 989803204341, 989803159571, 989803159581, 989803160241, 989803160251, 989803160261, 989803182921, 989803182931, 989803105531, 989803105541, 989803105561; UDI: N/A; System Codes 989803105531, 989803105541, 989803105561, 989803159581, 989803160241, 989803160251, 989803160261, 989803182921, 989803182931, 989803204301, 989803204311, 989803204321, 989803204331, 989803204341, 989803204511, 989803204521, 989803204531, 989803159571, 989803159581, 989803160241, 989803160251, 989803160261, 989803182921, 989803182931, 989803204301, 989803204311, 989803204321, 989803204331, 989803204341, 989803204511, 989803204521, 989803204531, M1920A, M1921A, M1923A.

View the original FDA recall notice

Why was this product recalled?

Disconnection of the adapter from the endotracheal tube can potentially be difficult or not possible.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America Llc take?

Consignees and Distributors were mailed an URGENT Medical Device Correction notice dated 4/7/25. Consignees were instructed to continue using the devices per the IFU and to ensure during setup that the airway adapter can easily be attached and detached from the breathing circuit/tubing before proceeding. The provided recall notification should be forwarded to those in the organization who use the device or where devices may have been transferred. Consignees are asked to utilize the provided QR code to complete a Response Form digitally; consignees can also email a hard copy Response Form to Recall.Response@fda.hhs.gov or by fax to 877-499-7723. Distributors are asked to forward the provided recall notification to customers no later than 5 days after notification of this recall event and to conduct a good faith effort to get a completed Response Form from each of their customers by conducting a minimum of three follow ups utilizing multiple contact methods. Consignees or Distributors with any questions are to contact 1-800-722-9377. Philips plans to issue an IFU addendum to address the recall.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology