Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit. Recall
Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1743-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 31, 2025
- FDA report date
- May 21, 2025
- Quantity in commerce
- 72 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit.
Affected model & UDI checker
Search the exact identifiers published for Z-1743-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| RPGH9472-01 | Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit. | Check source notice | Not specified |
| Not specified | Halyard RAPIDES E-PACK. Cardiovascular Procedure Convenience Kit. | Check source notice | 10809160211311 |
FDA code information
Model No. RPGH9472-01; UDI: 10809160211311; Lot Code: Lot: 1636640; Exp: 1/13/2028.
Why was this product recalled?
An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did AVID Medical, Inc. take?
An URGENT: MEDICAL DEVICE RECALL NOTIFICATION dated 3/31/25 was emailed to consignees. The notification indicates that this recall is due to Ethicon having recalled PDS II, PDS Plus, PROLENE, and ATHIBOND EXCEL sutures, which are components in the convenience kits being recalled by the recalling firm. Consignees are being instructed to discontinue use of the Johnson & Johnson Ethicon Sutures contained inside the recalled AVID kits identified in the recall notification and to add the provided labels to affected kits to ensure recalled sutures are discarded. Distributors are asked to forward the provided recall notification to their customers to which product was further distributed. Consignees are asked to return the provided response form via email. Questions regarding this recall event are to be directed to your distributor or sales representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1743-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology