Alinity m Resp-4-Plex Amplification (AMP) Kit (CE), List Number 09N79-090, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the qualitative detection and differentiation of RNA from influenza A virus (flu A), influenza B virus (flu B), Respiratory Syncytial Virus (RSV), and SARS-CoV-2 in nasopharyngeal (NP) swab specimens collected by a healthcare provider, from individuals with signs and symptoms of respiratory tract infection Recall
Alinity m Resp-4-Plex Amplification (AMP) Kit (CE), List Number 09N79-090, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the qualitative detection and differentiation of RNA from influenza A virus (flu A), influenza B virus (flu B), Respiratory Syncytial Virus (RSV), and SARS-CoV-2 in nasopharyngeal (NP) swab specimens collected by a healthcare provider, from individuals with signs and symptoms of respiratory tract infection
FDA recall Class II
Ongoing
FDA lists this Class II recall for Alinity m Resp-4-Plex Amplification (AMP) Kit (CE), List Number 09N79-090, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the qualitative detection and differentiation of RNA from influenza A virus (flu A), influenza B virus (flu B), Respiratory Syncytial Virus (RSV), and SARS-CoV-2 in nasopharyngeal (NP) swab specimens collected by a healthcare provider, from individuals with signs and symptoms of respiratory tract infection. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1748-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 28, 2025
- FDA report date
- May 21, 2025
- Quantity in commerce
- 1447 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Alinity m Resp-4-Plex Amplification (AMP) Kit (CE), List Number 09N79-090, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the qualitative detection and differentiation of RNA from influenza A virus (flu A), influenza B virus (flu B), Respiratory Syncytial Virus (RSV), and SARS-CoV-2 in nasopharyngeal (NP) swab specimens collected by a healthcare provider, from individuals with signs and symptoms of respiratory tract infection
Affected model & UDI checker
Search the exact identifiers published for Z-1748-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Alinity m Resp-4-Plex Amplification (AMP) Kit (CE), List Number 09N79-090, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the qualitative detection and differentiation of RNA from influenza A virus (flu A), influenza B virus (flu B), Respiratory Syncytial Virus (RSV), and SARS-CoV-2 in nasopharyngeal (NP) swab specimens collected by a healthcare provider, from individuals with signs and symptoms of respiratory tract infection | Check source notice | 00884999049338 |
FDA code information
UDI-DI: 00884999049338; Lot Numbers: 409383 410627 411921
Why was this product recalled?
There were reports of an increase in reactive negative controls and false positive results.
What actions did Abbott Molecular, Inc. take?
An Urgent Field Safety Notice/ Field Correction Recall notification letter dated 4/24/25 was sent to customers. Necessary Actions: " Discard inventory of the lot listed in the above table. Contact Abbott Customer Support for replacement of any unused kit(s). " If you have forwarded any kits of this lot to other laboratories, please inform them of this Urgent Field Safety Notice / Field Correction Recall and provide a copy of this letter. " Return the associated Customer Reply form. Please review this information with laboratory personnel and retain this communication for future reference. If you have any questions regarding this communication, please contact your local Abbott representative, or call Abbott Technical Support at 1-800-553-7042 Option 2. We apologize for any inconvenience this may have caused your laboratory.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1748-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026