Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1769-2026

Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252 Recall

Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252

FDA recall Class II

Ongoing

FDA lists this Class II recall for Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1769-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 5, 2026
FDA report date
Apr 15, 2026
Quantity in commerce
233
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

1. UDI-DI (GTIN) 00840338603623 Lot Numbers AZ0124352 AZ0224352 AZ0324352 AZ0424352 AZ0524352 2. UDI-DI (GTIN) 00840338603630 Lot Numbers AZ0624352 AZ0724352 AZ0824352 AZ0924352 AZ1024352 3. UDI-DI (GTIN) 00840338604033 Lot Numbers AZ0124351 AZ0224351 AZ0524351 AZ0724351 AZ0924351

View the original FDA recall notice

Why was this product recalled?

Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.

What actions did Tornier, Inc. take?

On March 5, 2026, the firm notified affected customers of the product issue. Customers were instructed to locate affected product in their inventory. Relevant personnel should be informed that the affected product must not be used with Tornier HRS Max products, as they are incompatible. Product return is not necessary, but customers may contact the recalling firm to arrange for return if the product issue impacts their needs.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.