See Luer Cap Set, MPC-130, set, administration, intravascular Recall
See Luer Cap Set, MPC-130, set, administration, intravascular
FDA recall Class II
Ongoing
FDA lists this Class II recall for See Luer Cap Set, MPC-130, set, administration, intravascular. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1774-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 17, 2026
- FDA report date
- Apr 15, 2026
- Quantity in commerce
- 26900 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
See Luer Cap Set, MPC-130, set, administration, intravascular
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
UDI +B144MPC1300/$$529105203895, Lot Number 20389
Why was this product recalled?
Molded Products, Inc. is voluntarily conducting a medical device recall of the MPC-130 See Luer Cap Set based on a confirmed complaint of the threaded sleeve not being engaged and becoming un-attached.
What actions did Molded Products Inc take?
Molded Products, Inc. issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 02/17/2026 via USPS First Class Mail. The notice explained the issue and requested the following: Please take the following actions: 1. Please inspect your inventory to determine if you have any of the affected lot. Quarantine the affected product and initiate a return to Molded Products, Inc. for credit. 2. Share this recall notification will all users of the product to ensure they are also aware of this recall. 3. Complete the attached Medical Device Recall Customer Response Form and return to the contact noted below whether or not you have any of the impacted material so that Molded Products may acknowledge your receipt of this notification and process your return and credit. For complaints, adverse events, or questions, contact Teresa Wooster, twooster@moldedproducts.com; Ph# 800-435-8957 ext. 236 Monday Friday 8:00am-4:30pm (CT), Fax# 800-227-7935.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1774-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology