True 3D Viewer, Model: EP-00003 Recall
True 3D Viewer, Model: EP-00003
FDA recall Class II
Ongoing
FDA lists this Class II recall for True 3D Viewer, Model: EP-00003. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1785-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 21, 2018
- FDA report date
- May 28, 2025
- Quantity in commerce
- 18
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
True 3D Viewer, Model: EP-00003
Affected model & UDI checker
Search the exact identifiers published for Z-1785-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| EP-00003 | True 3D Viewer, Model: EP-00003 | Check source notice | Not specified |
| Not specified | True 3D Viewer, Model: EP-00003 | Check source notice | 00851325007003 |
FDA code information
UDI-DI: 00851325007003. Full release software versions: 1.6.1.3080.428, released 13-Apr-2017 1.6.2.3670.435, released 07-Dec-2017 Develop Versions, for research use: 0.00.2695.422, released 15-Feb-2017 0.00.2927.423, released 15-Feb-2017 0.00.2964.423, released 24-Feb-2017 0.00.3102.428, released 04-Apr-2017 0.00.3245.428, released 05-May-2017 0.00.3302.431, released 02-Jun-2017 0.00.3344.432, released 18-Jul-2017 0.00.3412.433, released 14-Aug-2017 0.00.3656.442, released 09-Nov-2017 0.00.3669.442, released 05-Dec-2017 0.00.3683.445, released 02-Jan-2018 0.00.3702.447, released 11-Jan-2018 0.00.3730.447, released 08-Feb-2018
Why was this product recalled?
With certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image processing/review/analysis/communication/media interchange acquired from CT/MRI/ultrasound, and pre-operative analysis of surgical options), can result in incorrect display of image orientation, which could result in misdiagnosis or incorrect surgical treatment plan.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Echopixel, Inc. take?
On 2/21/2018, correction notices were emailed to customers who were asked to do the following: a) Affected software versions should not be used with Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM. b) If clinicians are utilizing Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM volume data sets, verify, retrospectively, any potential risk to health to patients and further determine if patients need to be reevaluated and report this information to the firm. c) Pass a copy of this notice to those staff members who need to be aware of the issue and maintain awareness for a three-month period. d)If devices were transferred to other departments or organizations, provide a copy of this notice to these consignees. Firm will contact customers to provide information about the timing and delivery mechanisms of the software upgrade. If you have questions, contact the firm at: 1-844-273-7766 x2
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1785-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology