BRAHMS KRYPTOR compact REACT, REF: 89986, used with BRAHMS PlGF Plus KRYPTOR, REF: 859075N, and run on the BRAHMS KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor Recall
B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the B¿R¿A¿H¿M¿S KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor
FDA recall Class II
Ongoing
FDA lists this Class II recall for BRAHMS KRYPTOR compact REACT, REF: 89986, used with BRAHMS PlGF Plus KRYPTOR, REF: 859075N, and run on the BRAHMS KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1795-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 12, 2026
- FDA report date
- Apr 15, 2026
- Quantity in commerce
- 5,524
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
BRAHMS KRYPTOR compact REACT, REF: 89986, used with BRAHMS PlGF Plus KRYPTOR, REF: 859075N, and run on the BRAHMS KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor
Affected model & UDI checker
Search the exact identifiers published for Z-1795-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 89986 | BRAHMS KRYPTOR compact REACT, REF: 89986, used with BRAHMS PlGF Plus KRYPTOR, REF: 859075N, and run on the BRAHMS KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor | Check source notice | Not specified |
| 859075N | BRAHMS KRYPTOR compact REACT, REF: 89986, used with BRAHMS PlGF Plus KRYPTOR, REF: 859075N, and run on the BRAHMS KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor | Check source notice | Not specified |
| Not specified | BRAHMS KRYPTOR compact REACT, REF: 89986, used with BRAHMS PlGF Plus KRYPTOR, REF: 859075N, and run on the BRAHMS KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor | Check source notice | 04260157638976 |
| Not specified | BRAHMS KRYPTOR compact REACT, REF: 89986, used with BRAHMS PlGF Plus KRYPTOR, REF: 859075N, and run on the BRAHMS KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor | Check source notice | 04260157632196 |
FDA code information
REF: 89986, UDI-DI: 04260157638976, Lots: 40102+R78, 40102+R79, 40102+R80. REF: 859075N, UDI-DI: 04260157632196, All Lots.
Why was this product recalled?
Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Brahms GmbH take?
On 2/27/2026, correction notices were sent to customers informing them: 1. Continue to perform Quality Control testing in accordance with the current Instructions for Use prior to reporting patient results. 2. If Quality Control results fall outside plus-minus 20% of target values follow the troubleshooting procedures described in the User Manual. 3. Forward this letter to those who need to be aware within your organization or to any organization to whom the potentially affected devices have been transferred. Keep this notification on file. For further questions, please contact the firm's customer service on 1-800-232-3342 or e-mail to: USFMTMGC-TechnicalService@thermo.com On 3/12/2026, Follow-up notices were sent to customers who were informed the root cause had been identified, which identified reaction plate affected devices. Customers were asked to do the following: 1) Stop using and dispose affected reaction plates. 2) Complete and return the response form. 3) Same as point 3 above.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1795-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology