META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690 Recall
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
FDA recall Class II
Ongoing
FDA lists this Class II recall for META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1806-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 18, 2025
- FDA report date
- May 28, 2025
- Quantity in commerce
- 75
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Affected model & UDI checker
Search the exact identifiers published for Z-1806-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 71642690 | META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690 | Check source notice | Not specified |
| Not specified | META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690 | Check source notice | 00885556132975 |
FDA code information
UDI-DI: 00885556132975. Lot: 24CT81426. Expiration: 25-Mar-2034
Why was this product recalled?
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
What actions did Smith & Nephew, Inc. take?
On 3/18/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inspect your inventory and locate any affected devices and quarantine them. 2) If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 3) Complete and return the notice response section via email to FieldActions@smith-nephew.com If you have any questions email the firm at the email address above.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1806-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology