Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1810-2025

Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C Recall

Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C

FDA recall Class I

Ongoing

FDA lists this Class I recall for Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1810-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Apr 22, 2025
FDA report date
Jun 11, 2025
Quantity in commerce
1853 total
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C

Affected model & UDI checker

Search the exact identifiers published for Z-1810-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349CCheck source notice10888277858060
FDA code information

UDI-DI 10888277858060 (EA); 40888277858061 (CASE) KIT LOTS 24CBJ925 24DBT405 24FBO318

View the original FDA recall notice

Why was this product recalled?

Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology