Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1810-2026

LinkBio CORE Workstation, a component of the CORE Shoulder System. Product Code: 87-9136. Recall

LinkBio CORE Workstation, a component of the CORE Shoulder System. Product Code: 87-9136.

FDA recall Class II

Ongoing

FDA lists this Class II recall for LinkBio CORE Workstation, a component of the CORE Shoulder System. Product Code: 87-9136.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1810-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 12, 2026
FDA report date
Apr 22, 2026
Quantity in commerce
5 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

LinkBio CORE Workstation, a component of the CORE Shoulder System. Product Code: 87-9136.

Affected model & UDI checker

Search the exact identifiers published for Z-1810-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedLinkBio CORE Workstation, a component of the CORE Shoulder System. Product Code: 87-9136.Check source notice00810022403099
FDA code information

Product Code: 87-9136. UDI-DI: 00810022403099. All serial numbers of the CORE Workstation are affected.

View the original FDA recall notice

Why was this product recalled?

The CORE Workstation may display an incorrect "Planning Date " when viewing cases on the "Start Case" menu screen.

What actions did Linkbio Corp. take?

LinkBio notified consignees on about 03/12/2026 via email. Consignees were instructed to take note of the potential inaccuracy of the "Planning Date," that no product was required to be returned, and to complete and return the Distributor Reply form. LinkBio Corp. is investigating the root cause of the time/date sync issue and will take appropriate action to ensure accurate dates are displayed for the user.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology