Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1812-2025

PhenoMATRIX Recall

PhenoMATRIX

FDA recall Class II

Ongoing

FDA lists this Class II recall for PhenoMATRIX. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1812-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 6, 2025
FDA report date
May 28, 2025
Quantity in commerce
14
Distribution
See key details
FDA cause
No Marketing Application

What is being recalled?

PhenoMATRIX

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

All software versions

View the original FDA recall notice

Why was this product recalled?

AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Copan WASP take?

On 5/6/2025, recall notices were mailed to customers informing them of the following: Ensure the plate images are reviewed against the actual agar plate to verify the assessment you performed by the digital image inspection. Complete and return the response form via email to regulatory.wasp@copangroup.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.