Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1814-2025

Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F Recall

Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F

FDA recall Class I

Ongoing

FDA lists this Class I recall for Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1814-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 1, 2025
FDA report date
Jun 11, 2025
Quantity in commerce
11,358
Distribution
See key details
FDA cause
Process control

What is being recalled?

Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F

Affected model & UDI checker

Search the exact identifiers published for Z-1814-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CPRM1116FAdult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116FCheck source noticeNot specified
FDA code information

UDI-DI: EA: 10884389164822 CS: 40884389164823 Lots: 526782 526790 526796 526797 526798 526800 526802 526804 526805 526806 526807 526808 526809 526810 526814 526815 526816 526811 526817 526818

View the original FDA recall notice

Why was this product recalled?

Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequences, including hypoxia, hypercapnia, organ failure, and death.

What actions did SunMed Holdings, LLC take?

Recall initiated May 1, 2025. Customers instructed to immediately cease use of any affected product and quarantine all affected product. Customers were instructed to return or destroy affected product. AirLife will send new replacement product to customers once disposition is confirmed. If you need replacement products to be sent to you urgently, please call AirLife directly on 1-800-433-2797 and the firm will make every effort to accommodate your needs.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology