ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Recall
ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.
FDA recall Class II
Ongoing
FDA lists this Class II recall for ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1817-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 22, 2023
- FDA report date
- Jun 4, 2025
- Quantity in commerce
- 23 units
- Distribution
- See key details
- FDA cause
- Labeling Change Control
What is being recalled?
ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.
Affected model & UDI checker
Search the exact identifiers published for Z-1817-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| HC-13120-03M | ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. | Check source notice | Not specified |
FDA code information
Model Number: HC-13120-03M. UDI-DI: B278HC1312003M0. Serial Numbers: P230274-1XX (XX represents 23 parts in sequential order from 101-123).
Why was this product recalled?
Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.
What actions did Onkos Surgical, Inc. take?
ONKOS Surgical notified their distributors on about 12/22/2023 via email. They were instructed to segregate and quarantine the product immediately, complete and return the "Product and Distribution Information" form within 2 business days and return any existing implants on hand within 2 business days after signing the acknowledgement. They were also instructed that they will receive replacement implants and to place them in their modular segmental collar stem implant packs.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1817-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology